STERILISATION
Syllabus:
General introduction to the microbes and effect of temperature on the
growth of Microorganisms.
Principles and processes of
sterilisation, sterilisation by filtration techniques and mechanisms involved
in these techniques.
Questions:
1. What are the different methods of sterilization?
Discuss in detail about two suitable examples. (1992) [4+12]
2. Mention suitable methods for the sterilization of
thermolabile drugs and antibiotics. (93) [8]
3. Write a note on thermal methods of sterilization.
give their advantages and limitations. (96)
[16]
4. Define sterilization. Name different methods of
sterilization with principle involved for sterilization of thermolabile liquids
in pharmaceutical industry. (98) [16]
5. What is sterilization by filtration? (98) [8]
INTRODUCTION
Among the many facets
of patient care is that of the sterilization of a range of materials. These
include bacteriologically contaminated glassware and Petri dishes, dressings,
sutures, ligatures, surgical instruments, etc., as well as certain raw
materials and forms of pharmaceutical dosage. It is considered necessary to
sterilise all of these as they could constitute a potential health hazard to
patients.
Sterilization is the
process of killing or removing microorganisms. A sterile material is one that
contains no living organisms at all and the term sterile is therefore an absolute one. However, the killing or
removing of microorganisms follows a probability function, there always being a
finite chance of leaving some live microorganisms intact in the system so that
sterilization process are designed to minimize the probability of leaving
survivors.
However, with all
articles to be sterilized there is the chance that the sterilizing treatment
will have a detrimental effect. This is particularly true of pharmaceutical
dosage forms where it is important that the chosen process should not cause
changes in the formulation, which would reduce its therapeutic efficacy or
patient acceptability. For this reason, with the design of all sterilization
processes a balance has to be achieved between the maximum acceptable risk of
failing to achieve sterility and the maximum permissible concomitant damage
caused to the treated articles.
The methods currently
used to kill microorganisms involve exposing the preparation to an inimical
physical agent or condition for a known period of time. The agents used are of
three different types, namely elevated temperature, ionizing or UV radiation,
and toxic gases. The fourth way of achieving sterility is restricted to
solutions and gases and involves passing the solution through a filter that
will retain the microorganisms.
DEFINITIONS
1.
Antiseptic: A substance
that arrests or prevents the growth of microorganisms by inhibiting their
activity without necessarily destroying them.
Can be applied on human body.
2.
Bactericide: Anything that
kills bacteria.
3.
Bacteriostat: Anything that
arrests or retards the growth of bacteria.
4.
Disinfection: A process that
removes infection potential by destroying microorganisms but not ordinarily
bacterial spores. Disinfectants are
generally meant for application on inanimate objects.
5.
Germicides: A substance that
kills disease microorganisms (i.e. pathogens / germs) but not necessarily
bacterial spores.
6.
Sterility: The absence of
viable organisms.
7.
Sterilization: A process by
which all viable forms of microorganisms are removed or destroyed.
8.
Viable = growing bacteria
(or microorganisms) + spores
9.
Vegetative
microorganisms = growing
organism
DEFINITION OF STERILISATION
Sterilization is a process whereby all the viable life forms
are either killed or removed (by filtration) from a product.
‘Sterilisation’ and ‘Sterility’ are two absolute terms i.e. a sterile product
should be absolutely free from viable (vegetative + spores) microorganisms.
METHODS OF STERILISATION
STERILISATION METHODS
PHYSICAL CHEMICAL MECHANICAL
with
ethylene oxide bacteria
proof
Dry heat
|
Moist heat
|
Radiation
|
||
i)Heating in hot air oven
ii)Flaming
iii)Infra-red
radiation
|
i) Moist heat in autooclave
ii) Heating with a bactericide
iii) Heating at 1000C
iv)Pasteurization
v)Tyndalization or Fractional sterilisation.
vi) Heating at 55 to 600C
|
i) By exposure to UV - rays
ii) By ionising radiation. (i.e gamma rays)
|
i) Berkefeld filter (could be porcelain, keiselguhr)
ii)Sietz filter (Asbestos pads)
iii) Sintered glass filter
iv) Sintered metallic filter
v) Membrane filters
|
DRY HEAT STERILIZATION
Principle:
The killing of microorganisms by heat is a function of the
time-temperature combination used. If the temperature is increased then the
time required for killing all the bacteria will be decreased.
Condition: Cycles recommended as per BP 1988 are:
A minimum of 1800C for
not less than 30 minutes.
A minimum
of 1700C for not less than 1 hour.
A minimum
of 1600C for not less than 2 minutes.
Mechanism of killing the bacteria:
The vital constituents of cells such as proteins (enzymes) and nucleic
acids are denatured by oxidation.
Biological indicator: Spores
of Bacillus subtilis var. niger. or,
Spores
of Bacillus globigii or,
Spores
of Clostridium tetani (atoxigenic)
Application of dry heat sterilization:
Dry heat is used to sterilize
1.
Glass ware e.g. test tubes,
Pasteur-pippettes, petridishes, flasks, glass syringes etc. The glass wares
should be prewashed with apyrogenic water.
2.
Porcelain and metal equipment such as forceps,
scalpels, scissors etc.
3.
Dry materials in sealed
container like powders.
4.
Fats, oils and greasy materials (like petroleum
jelly) those are impermeable to moisture.
Dry heat sterilization is not used in the following cases:
1.
Aqueous solution cannot be
sterilized by dry heat beacause the water will evaporate if the container is
kept open. If the container is kept close then it may burst due to pressure
developed by the steam.
2.
Surgical
dressings cannot be sterilized by dry heat because cotton or other cellulosic
material will get chared at the high temperature.
3.
Powders that cannot withstand the high temperature
are not sterilised by dry heat method.
HEATING IN A HOT AIR OVEN
·
A hot air oven is a double walled chamber made up of
mild steel or aluminium. The door is insulated with asbestos gasket to minimize
the heat loss. To reduce heat loss by conduction insulating material (glass
fibre or puff) is used in between two walls.
·
The inner surface is made reflecting to reduce heat
loss by radiation.
·
It consists of one or more shelves.
·
It is thermostatically controlled. Heaters are
fitted to heat the air inside. The heat is circulated by normal convection. To
spread the heat uniformly forced convection is arranged by fitting fans in
strategic places.
·
For effective utilization of the oven and to obtain
correct results it should be uniformly packed or loaded so that all the objects
are exposed to sterilization temperature for the required period of time.
·
Care should be taken so that once in operation the
oven is not opened in the middle of the cycle.
FLAMING
This is an emergency
method, the forceps-tips, the surfaces of the scalpels and the needles may be
sterilized by holding the items directly in the flame of a Bunsen burner. This
method is generally done in microbiology.
INFRARED RADIATION
·
Infrared radiation (IR) is a thermal radiation, i.e.
when absorbed by some article its energy is converted to heat and therefore it
is often known as radiant energy.
·
A tunnel having an IR source is used for this
purpose. The instruments and glass wares are kept in trays are passed through
this tunnel keeping on the conveyor belt, at a controlled speed exposing them
to a temperature of 1800C for 17 minutes, thereby achieving the
sterility. This is a continuous process and is used in hospitals for regular
supply of sterile syringes and other apparatus.
·
Heating at or above 2000C by IR in vacuum
is employed as a means of sterilizing surgical instruments.
·
Cooling is hastened, (after the heating cycle)
during the cooling period, by admitting filtered N2 to the chamber.
MOIST HEAT STERILIZATION
Principle:
Mechanism of killing of microorganisms:
Bacterial death by
moist heat is due to denaturation and coagulation of essential protein
molecules (enzymes) and cell constituents.
Conditions:
The USP XXI and BP 1988 recommended the following condition:
·
Pressure: 15
lb / square inch (psi)
·
Temperature: 1210C
·
Time: 15
minutes
·
The following combinations of temperature and
holding time are normally employed for sterilizing by heat in autoclave:
Holding temperature (0C)
|
Holding time (min)
|
115 to 118
121 to 124
126 to 129
134 to 138
|
30
15
10
3
|
Biological indicator: Spores
of Bacillus stearothermophilus or
Spores
of Clostridium sporogenes.
Principles of sterilization by steam under pressure:
Pressure itself has no
sterilizing power. Steam is used under pressure as a means of achieving an
elevated temperature.
Steam production: This may be
achieved in two ways:
1.
In a small scale, steam may be generated from water
within the sterilizer and since water is present this steam is known as ‘wet
saturated steam’.
2.
For large scale sterilizers dry saturated steam may
be piped from a separate boiler.
Penetration of steam:
Steam flows quickly
into every article in the load ( and into porous articles). This is due to its
condensation creating a low-pressure region into which more steam flows.
Rapid heating: The saturated
steam heats the load rapidly due to the release of its considerable amount of
latent heat.
Moist heat: The condensate
produced on cooling hydrates the microorganisms and thus helps in coagulating
microbial proteins.
No residual toxicity: The product is
free from toxic contamination.
The design and operation of autoclave:
A portable autoclave
is an elaborate pressure cooker. It is a hollow cylindrical vessel fitted with
a lid that can be tightly secured in a position by nut-bolts or screws. The
body is made up of aluminium or steel or gun metal.
The lid is provided
with a steam vent, a safety valve and a pressure or temperature gauge. It is
heated electrically or gas operated. The electrical element is fitted at the
bottom of the autoclave. First water is added so that the heating element is
fully immersed in water. Then the materials to be sterilized are placed over a
perforated platform.
Heater is switched on.
Initially the steam-vent (outlet) is opened. The displaced air is first
removed, then the vent is closed. The pressure will then rise to 15 lb/sq inch
(psi) and as a result the temperature inside will rise to 1210C. At
this condition the autoclave is kept for 15 minutes, then the heater is
switched off. When the pressure inside and outside the autoclave equals, the
steam-vent is opened and the lid is removed to take out the sterilized
articles.
N.B. Precaution to prevent sterilization failure by
autoclave:
1) Incomplete
evacuation of air from the sterilizing chamber:
·
Air occupies the space in the
sterilizer before steam is generated or admitted. Air in the sterilizer forms a
thin layer that adheres to every surface on which steam condensation occurs.
Since air is poor conductor of heat it makes a barrier to heat transfer.
·
The
temperature of a mixture of air and steam under a given pressure is less than
that of a pure steam alone. This means that although the autoclave is kept at
the desired pressure the temperature may not be sufficient to provide sterilization.
·
Time
required for attaining the autoclaving temperature depends on the type of load
(materials to be sterilized) and their arrangement inside the autoclave. Hence,
practically, the holding temperature is held for more time than the prescribed one.
Application of autoclave:
It
is used to sterilize anything, which is not injured by steam and high
temperature of sterilization. These includes:
1.
Aqueous parenteral solutions e.g. distilled water,
saline solutions.
2.
Aqueous liquid media e.g. liquid media with or
without carbohydrate and gelatin.
3.
Surgical dressings and fabrics.
4.
Plastic and rubber closures.
5.
Metal instruments.
6.
Glass apparatus and containers.
HEATING WITH A BACTERICIDE
This method is used
for sterilizing aqueous solutions that are thermolabile to withstand normal
autoclaving temperature.
Condition:
Temperature: 98 to 1000C boiling
water
Time: 30 minutes
Bactericides: Bactericide that is compatible
with the product, container
and closure:
e.g. for
injection
Chlorocresol 0.2%(w/v)
Phenylmercuric nitrate (PMN) 0.002%(w/v)
Phenylmercuric acetate (PMA) 0.002%(w/v)
e.g. for eye drop
Thiomersal 0.01%(w/v)
Chlorhexidine acetate 0.01%(w/v)
PMA and PMN 0.002%(w/v)
Benzalkonium chloride (BAC) 0.01%(w/v)
Applications:
In the following cases this
method can be used:
1.
Injection fluids in the final container (i.e.
terminal sterilization).
2.
Eye drops.
**In the following cases
this method cannot be used:
1.
Solution of medicaments intended for intrathecal,
intra-atrial, peridural, intra-peritoneal, intra-cisternal and any other route
of injection giving access to the cerebro-spinal fluid and for intracardiac and
intra-ocularly.
2.
Large volume parenteral fluids having a volume
greater than 15 ml is not recommended to use bactericide because, though the
concentration of the bactericide in the solution is less but the total amount
entering into the system may be considerable due to large volume. e.g.
bactericides are not used in saline solutions.
N.B.
·
The
use of combination of heat and antimicrobial agents is synergistic and provided
the time of exposure is sufficient, permits the use of a lower sterilizing
temperature.
·
An
interesting modification of heat-chemical synergism is found if the formulation
is at high or low pH and heat on bacteria enables sterilization to be achieved
at reduced temperature. e.g. Phenobarbital injection B.P.
·
It is
possible also that a medicament possess intrinsic antibacterial activity which
again may permit sterilization at the lower temperature.
TYNDALISATION or FRACTIONAL STERILIZATION
Instrument used : Arnold’s sterilizer
Condition: Arnold’s
sterilizer employs streaming steam at a temperature of approximately 1000C for a period of 20 minutes for three consecutive days.
e.g. Agar media is sterilized by this method.
Principle:
In this process it is
presumed that on the first day all the vegetative cells will be killed and some
spores may germinate in the following day. So the second day’s sterilization
cycle will kill the vegetative cells. Similarly some spore may still remain.
Therefore for an additional 24 hours period is allowed to elapse to make sure
all ‘spores’ have germinated into vegetative cells.
N.B.

Causes of failure in this process:
It may be seen
unless the spores are germinated the method will fail to sterilize. Failure may
be due to the following reasons:-
1.
The
media may be unsuitable for the germination of the spores, e.g. distilled water
is not a favorable environment for the growth of bacteria and will not permit
spores to germinate into vegetative cells.
2.
Spores
of anaerobic bacteria, if present, will not germinate in a media in contact
with air.
3.
If any
preservatives are present in the product it may not allow to germinate the
spores.
4.
If a
product has intrinsic antimicrobial properties then it will not allow to spore
germination.
Application of Arnold’s
sterilizer (Tyndallization)
The Arnold’s
sterilizer is used principally for the sterilization of gelatin, milk and
carbohydrate media. Higher temperature in the autoclave, larger single exposure
in the Arnold’s sterilizer may hydrolyze or decompose carbohydrates and prevent
gelatin from solidifying. Obviously such media would then be unsuitable for
use.
PASTEURIZATION
Pasteur developed the
procedure of gentle heating (Pasteurization) to prevent spoilage of beers,
wines by undesired contaminating microbes. This process has no pharmaceutical
application. This is used for processing of milk in order to kill the
pathogenic bacteria without affecting the taste and nutritional value of the
milk. Following two methods are used:
Holding method:
The milk is heated in
tanks at 620C for 30 mins
while it is gently stirred and the steam is flown on the surface to disperse
the foam.
High temperature short time method (H.T.S.T. / Flash method):
In this process the
milk is rapidly raised to 71.60C
held at this temperature for at least 20
seconds and quickly cooled.
The common milk born
pathogens (Mycobacterium tuberculosis,
M. bovis, Brucella abortes) do not
form spores and are reliably sterilized by this procedure. In addition to this,
total bacterial count is generally reduced by 97 - 99%. therefore in true sense this technique is not an absolute
sterilization process.
HEATING AT 55 TO 600C
Certain vaccines are
sterilized by this method, the vaccines, in sealed containers, are immersed in
the water bath which is thermostatically controlled. They are heated at 55 to
600C for 15 minutes to 60 minutes depending upon the nature of
organisms.
This method will kill bacteria but will not destroy its antigenicity.
*Maximum care should be taken while processing the vaccines so that
they are not contaminated by spores which cannot be destroyed at this
temperature.
Advantages of moist heat
sterilization:
1.
High heat content plus rapid heat transfer.
2.
Destroys micro-organism more efficiently than dry
heat.
3.
It can be used for a large number of injections,
ophthalmic solutions, irrigants, dialysis fluids etc.
4.
It rapidly penetrates porous materials and is
therefore very suitable for sterilizing surgical dressings and materials.
5.
The process is adaptable for plastic containers and
some other special dosage forms.
6.
It is more suitable than dry heat for sharp
instruments.
7.
Accurate control and monitoring of the process is
possible.
8.
No toxic contaminants are left in the materials
sterilized.
Disadvantages of moist heat
sterilization:
1.
It is not suitable for anhydrous materials such as
powders and oils.
2.
It cannot be used for thermolabile substances.
3.
It does not destroy pyrogens.
GASEOUS
STERILIZATION
This process involves
exposure of materials to sterilize gasses such as ethylene oxide, formaldehyde, glutaraldehyde, propylene oxide.
ETHYLENE OXIDE
Ethylene oxide is the
only gas that is successfully used on a large scale of industrial and medical
applications.
Structure:

Mechanism of action: Ethylene oxide acts by alkylation.
Biological indicator: Bacillus subtilis var. niger.
Factors affecting
sterilization:
Sterilization efficiency is influenced by
1.
the concentration of ethylene-oxide
2.
the humidity of the sterilizing atmosphere
3.
the temperature of sterilization
4.
time of exposure
5.
physical nature and permeability of the load.
6.
atmospheric preconditioning of the load before
sterilization.
N.B. Organisms
treated in a dried state are several fold more resistant than when preconditioned
in an atmosphere of high relative humidity before treatment, almost regardless
of the relative humidity of the sterilizing environment.
With gases,
certain minimum concentration is necessary as otherwise exposure time becomes
unnecessarily prolonged. At temperatures of 40 to 500C, ethylene
oxide concentrations in excess of 400 mg/L are necessary.
With a sterilizing
environment humidity in the range of 30 to 60%, the load having being
preconditioned for 24 hours at a relative humidity in excess of 60%.
Inactivation factors of the order of 1017 for Bacillus subtilis spores on non-hygroscopic surfaces can be
achieved in about 2 hours. This is assuming rapid and complete penetration of
the load preferably aided by the evacuation of the sterilizing chamber before
treatment.
Because of the
uncertainty of the efficiency of the process, exposure time of 3 to 4 hours in
the above conditions are more usual to allow a wide safety margin.
Condition:
Relative humidity (rh) 33%
Temperature 250C
[generally 40 - 500C, maximum limit 600C
Concentration
of ethylene oxide
|
Time (Temp. 540C)
|
900 mg / litre
450 mg / litre
|
3 hour
5 hour
|
Precautions
·
It is important to ensure that the articles for
gaseous sterilization are scrupulously clean. Organic material reduces the
efficiency of the process.
·
After sterilization ethylene-oxide gets strongly
adsorbed by a wide variety of substances. Desorption of the gas is done by
airing the materials in a well-ventilated room for sometime or a powerful
vacuum is applied.
·
Ethylene oxide sterilizers:
The features of a suitable equipment include:
1.
An exposure chamber that is gas tight.
2.
A means of heating the chamber: e.g. a steam or hot
water jacket or, heating element clipped to the outside of the chamber.
3.
A system for adding water to provide the right
humidity.
4.
A means of extracting air before and after the
sterilization.
Use:
Ethylene oxide possesses the ability to penetrate paper, a number of
plastics and rubber. So disposable syringes, hypodermic needles, prepackaged
materials etc. are sterilized by this method.
Advantages:
1.
It is suitable for thermolabile substances, because
it can be carried out at room temperature or only slightly above.
2.
It does not damage moisture-sensitive substances and
equipment because only a low humidity is required.
3.
It can be used for prepackaged articles, because of
the great penetrating power of ethylene oxide.
4.
Though ethylene oxide is highly reactive compound
comparatively few materials are damaged by this process.
Disadvantages:
1.
It is slow. Long exposures and desorption periods
are necessary and therefore cannot be used in emergency.
2.
The running costs are high.
3.
Ethylene oxide is inflammable. To overcome it
ethylene oxide is mixed with inert gases such as carbon dioxide, fluorinated
hydrocarbons.
Some marketed mixtures are:
Cryoxide : 11%w/w ethylene oxide
79% w/w
trichlorofluoromethane
10% w/w
dichlorodifluoromethane
Sterethox 12% ethylene oxide
8%
dichlorodifluoromethane
4.
Toxic substances, such as ethylene chlorohydrin are
produced in some materials.
FORMALDEHYDE
Like ethylene oxide this is an alkylating agent but it is generally
inferior for use as a sterilizing agent. Because
·
formaldehyde has poor penetrating power and
is readily inactivated by organic matter.
·
High concentrations are difficult to maintain in the
atmosphere because it tends to deposit in the form of solid polymers on contact
with cool surfaces.
Condition:
Input concentration: 2
g/L of HCHO in subatmospheric steam
Temperature: 900C
Exposure time: 3 hours
Formaldehyde can be obtained as :
1.
Formaldehyde solution (Formalin) B.P.
Approximately 37%w/w
containing stabilizers to prevent deposition of solid polymers
2.
Tablets of paraformaldehyde
* In case of formalin addition of KMnO4 produces heat by
oxidation and the gas will be vaporized.
Use: Dressing packs.
Disadvantages:
1.
Low penetrability. It cannot penetrate polymeric
packaging.
2.
Polymerize to inactive forms on the surface of low
exposure temperature.
3.
Can fall in the active concentration
4.
Cannot sterilize narrow lumen.
Solution of the above problems:
1.
Reducing the size of the load.
2.
Increasing exposure temperature.
STERILIZATION BY RADIATION
Radiation can be divided into two groups:
1. Electromagnetic waves: (i) infra-red radiation (IR) (iii) X-rays
(ii) ultraviolet radiation (UV) (iv) gamma
rays
2. Streams
of particulate matter (i) alpha radiation (ii) beta
radiation
For sterilization infrared, ultraviolet, gamma radiation and high
velocity electrons (a type of beta radiation) are used for sterilization.
Radiation
|
Wave length
|
Energy
|
UV-radiation
Gamma radiation
High velocity electrons
|
190 to 370 nm
1 to 10-4 nm
|
5 eV
1.3 MeV*
4MeV
|
* MeV = Million electron volt
ULTRAVIOLET RADIATION
Source: Low current of high voltage
is passed through mercury vapor in an evacuated tube made of borosilicate
glass.
Dose of sterilizing radiation: 10 to 60
microwatts / cm2 reduce the populations of vegetative cells by 90%
in a short period.
Mode of action: Only a narrow range of
wavelength (220 to 280 nm) is effective in killing micro-organisms, and
wavelengths close to 253.7 nm are the most effective. It has been found that a
maximum biological efficiency exists at 253.7 nm, which is also the absorption
peak of isolated DNA, this suggests strongly therefore that DNA or the nucleic
acid is the target for UV-induced lethal events.
N.B.
The absorption and subsequent reactions are predominantly in the pyrimidines
of the nucleic acid. One important alteration is the formation of a pyrimidine
dimer in which two adjacent pyrimidines become bonded. Unless dimers are
removed by specific intracellular enzymes, DNA replication can be inhibited and
mutation can result.
Use: (i) Surface
sterilization
(ii) Sterilization
of clean air and water in thin layers.
Disadvantages:
1. It has very poor penetration power.
[N.B. UV radiation does not penetrate normal packaging
material, such as glass or plastic and hence, it is not used to sterilize
pharmaceutical dosage forms.]
2. UV-light sterilization is not absolutely
reliable because DNA may get repaired in some favorable condition.
(i) Photo reactivation:
If a suspension of bacteria is exposed to UV light to several minutes only a
small fraction will be able to form colonies in a nutrient media and thus
appear to be viable. However, if a sample of the irradiated cells is then
exposed to visible light for several minutes the fraction of survivors in this
sample will be much higher due to photo-reactivation of some of the damaged
cells.
Photo-reactivation never reaches 100% efficiency; i.e. not
all cells recover. Photo reactivation is actually caused by a light activated
enzyme that recognizes thiamine dimers in the DNA and cleaves them so
that normal DNA structure can be stored.
(ii) Excision repair or
Dark repair: Some microorganisms possess a second mechanism for repairing
damage to DNA caused by UV radiation; this mechanism is not light dependent and
two enzymes are involved, a dimer specific endonuclease and a dimer
specific exonuclease. Exonuclease excises the dimer together with a
number of neighbouring nucleotides. The missing piece of DNA strand resulting
from this excision is repaired by enzymatic insertion of nucleotides
complementary to the good strand. Two enzymes accomplish this job - a DNA
polymerase synthesizes the required segment and a DNA ligase
reestablishes its link in the strand.]
3. Bactericidal UV -light causes eye problems
and erythema (reddening) of the skin if those parts are exposed to UV-
radiation for a prolonged period of time.
IONISING
RADIATION
Mode
of action of ionizing radiation:
Ionizing radiation can cause excitations, ionization and
where water is present free radical formation. Free radicals are powerful
oxidizing ( OH, HO2) and reducing (H) agents, which are capable of
damaging essential molecules (enzymes and DNAs) in living cells. This results
in cell death.
Biological
indicator: Bacillus pumilis.
HIGH SPEED RADIATION
Source: This type of
sterilizing radiation is most widely used in Denmark and the USA. In a machine
known as a van de Graph accelerator electrons are generated from a suitable
source and then accelerated along a highly evacuated tube by a tremendous
potential difference between the ends.
Dose: 5 MeV to 10 MeV
(Million electron Volt).
Use: The beam, which
is narrow and intense, is used to irradiate articles on a conveyor-belt.
GAMMA RAYS
Source: Radiation from
the radioactive isotope of Cobalt 60 Co, is used as a source of
gamma emission.
Dose: 1.25 MeV,
Some users take the adequate dose as 2.5 Mrad (Mega
radiation unit)
Procedure: Articles for
sterilization by radiation are packed in boxes of standard size, which are
suspended from a monorail and sterilized by slow-passage around the gamma-ray
source.
Uses:
·
Articles regularly sterilized on a commercial scale
include plastic syringes, catheters, hypodermic needles and scalpel blades,
adhesive dressings, single-application capsules of eye-ointment and catgut
·
Containers made of polyethylene and packaging
materials using aluminum foil and plastic films.
Disadvantages:
·
The damage of the cells is mediated through
radiation-induced free radicals in water, hence the extent of degradation is
found to be maximum in pharmaceuticals in aqueous solutions. Practically the
amount of degradation in aqueous solution is so great that this sterilization
method is only confined to sterilize surgical sutures, instruments etc.
·
The necessary apparatus is much expensive for
installation in hospital. It is employed commercially for the sterilization of
large amount prepackaged disposable items such as plastic syringes and
catheters, which are unable to withstand heat.
STERILIZATION BY FILTRATION
This method is used for sterilizing thermolabile
solutions, which will otherwise be degraded by other conventional heating methods.
The drug solutions are passed through the sterile
bacteria proof filter unit and subsequently transferring the product
aseptically into the sterile containers which are then sealed.
The process involves considerable hazards. Hence IP and
BP require that the tests for sterility be carried out on the filtered product.
Biological
indicator: Micromonospora diminuta.
Procedure:
The solutions to be sterilized is
passed through the filter and collected in the sterile receiver by the
application of positive pressure to the nonsterile compartment or negative
pressure to the sterile side.
[Precautions:
1. Precautions should be taken
to avoid excessive positive or negative pressure.
2. Prolonged filtration must
be avoided to prevent the growth of a contaminant through the filter media and
entry into the sterile solution.
3. Filter elements of
porcelain or of sintered materials those are used repeatedly should be tested
for cracks or leaks prior to each use.
4. Fiber
shedding filters such as asbestos filters should be avoided. If, however, the
use of this type of filters is unavoidable, a fiber-retaining filter should be
used down stream of the asbestos filter.
5. The effective pore size of
the sterilizing filters should be confirmed before use and the continued
integrity of the filters should be confirmed after use.
6. The process of
sterilization must be validated by monitoring the microbial load in the
solution to be filtered.
7. For filters of a
sintered or fritted construction a procedure such as bubble-point test or
diffusion rate test should be carried out in accordance with the manufacturer's
recommendation. The integrity of any new or modified filtration system should
be determined before it is placed in service.]
Advantages
of sterilization by filtration:
1. Thermolabile solutions can be
sterilized.
2. It removes all the living
microorganisms.
Disadvantages
of sterilization by filtration:
1. Filters may break down
suddenly or gradually on use.
2. Sterility testing is
obligatory on the filtered solution.
3. Filter media may be absorbed
on the filter surface.
4. Viruses are not removed by
filtration.
5. Suspensions and oils cannot
be sterilized by this method due to their heavy load of particulate
matters and viscosity.
Mode
of action:
The filters are thought to
function by one or usually a combination of the following:
1. Sieving or screening,
2. Entrapment,
3. Electrostatic attraction.
When a particle is larger than the pore size of the
filter the particle is retained on the filter - this known as sieving or screening.
Entrapment
occurs when a particle smaller than the size of the pore enters into the pore
channel and lodges onto the curves of the channel while passing through it.
Electrostatic attraction causes
particles, opposite in charge to that of the surface of the filter pore, to be
held or adsorbed onto the surface.
MEMBRANE FILTERS
·
Membrane filters are made of cellulose-derivative
(acetate or nitrate). They are very fine. They are fixed in some suitable
holders.
·
Nominal pore size
is 0. 22 ± 0.
02 mm or less is required.
·
The membranes are brittle when dry. In this
condition they can be stored for years together. They become very tough when
dipped in water.
·
They are sterilized by autoclaving or by ethylene
oxide gas. They cannot be sterilized by dry heat as they decompose above 1200 C.
·
They are suitable for sterilizing aqueous and oily
solutions but not for organic solvents such as alcohol, chloroform etc.
·
Membrane filters are generally blocked by dirt
particles and organisms. Pre-filtration (through glass-fibre paper prefilter)
reduces the risks of blockage of the final filter.
Examples
of membrane filters:
MF-Millipore – it is a mixture of cellulose esters
Sartorius Regular – it is made of cellulose nitrate
Gelmen Triacetate Metricel – cellulose triacetate
SINTERED (or FRITTED) GLASS
FILTERS
![]() |
Borosilicate glass is finely powdered in a ball-mill and
the particles of required size are separated. This is packed into disc mounted
and heated till the particles get fused. The disc thus made have pore size of 2
mm and are used for filtration.
They are cleaned with the
help of sulfuric acid.
SIETZ FILTER: They are made of asbestos pad.
[For further details see Cooper & Gunn Dispensing
pp. 582]
CERAMIC FILTERS: They are
made of either porcelain or keiselghur. These are supplied in the shape of
candles mounted to metallic joint.
TESTING OF FILTERS:
The BP requires that the
integrity of an assembled sterilizing filter be verified before use and
confirmed after use by means of a suitable test.
Bacteriological
test:
A diluted solution of broth culture of Serratia marcescens is passed though the
filter and the filtrate is collected aseptically and incubated at 250C
for 5 days. The filter passes this test if no growth is found after 5 days of
incubation.
Bubble
point test:
The bubble point of a test filter is the pressure at which the largest pore of a
wetted filter is able to pass air.
Objectives:
1. Filtration should normally be
performed at pressures lower than the bubble point of a membrane. This prevents
gas from passing through the filter at the end of a filtration cycle and
thereby prevents excessive foaming.
2. For testing membrane
efficiency and integrity.
Theory:
Membrane filters, which have discrete uniform passages
that penetrate from one side of the media to the other, can be regarded as
fine, uniform capillaries. The bubble point test is based on the fact that when
these capillaries are full of liquid, the liquid is held by surface tension.
The capillary pressure is higher in the case of a small pore
than in that of a large pore. The same is true for pores in a membrane. The
bubble point pressure is governed by the following equation:
Where, P = bubble point pressure,
K = shape correction factor (experimental
constant),
D = pore diameter,
g = Surface- tension of the
liquid
q = Liquid-to-membrane
contact angle (angle of wetting)
Procedure:
1. First the membrane is wetted
and usually has a liquid above and a gas below.
2. Since the pore is full of
liquid there is no passage of gas at zero pressure. The pressure is increased
gradually. At bubble point pressure a small bubble will form at the largest
opening.
3. As the pressure is further
increased, rapid bubbling begins to occur.
Inference:
Bubble point pressure for a given set of membrane and
liquid is constant. If the bubble point pressure got reduced it can be
concluded that the membrane integrity is lost or efficiency is reduced.
MANUFACTURE OF STERILE BULK
POWDER
Some drugs are difficult to sterilize terminally (i.e.
when the total product is ready within the container and package), the raw
materials (i.e. bulk drugs) of those products are required to be sterile.
Generally, the manufacture
of a sterile bulk substance usually includes the following steps:
1. Conversion of
non-sterile drug substance to the sterile form by dissolving in an (organic or
aqueous) solvent, sterilization of the
solution by filtration and collection in a sterilized reactor (crystallizer).
2. Aseptic
precipitation or crystallization of the sterile drug substance in the sterile
reactor.
3. Aseptic isolation
of the sterile substance by centrifugation or filtration.
4. Aseptic drying
(spray drying, lyophilization), milling
and blending of the sterile substance.
All the above mentioned
operations should be performed in closed systems, with minimal operator
handling.
Sterilization
of the processing equipment
Equipment used in the processing of sterile bulk
substances should be sterile and capable of being sterilized. This includes the
crystallizer, centrifuge and dryer.
A. Steam sterilization:
The method of choice for the sterilization of equipment and transfer lines
(pipings) is saturated, clean steam under pressure.
[In the validation of the sterilization
process for the equipment and of transfer line include the use of Biological indicators as well as Temperature sensors (e.g. Thermocouple
or Resistance Thermal Device) should be strategically located in cold spots
where condensate may accumulate. These include the points of steam injection
(inlet) and steam discharge (outlet) as well as in the cold spots.]
B. Formaldehyde:
Formaldehyde is used rarely due to its residue present in the environment and
the equipment after the sterilization process. This residue is very difficult
to remove.
SUMMARY OF INDUSTRIAL
STERILIZATION
Sterilization
process / Conditions
|
Sterilizable
articles
|
Nonsterilizable
articles
|
Moist
heat or Autoclave
15 lb/sqinch, 1210C, 15 mins
|
Anything heat stable,
i) Most types of solid and liquid media with
or without carbohydrate, gelatin
ii) Distilled water, saline solution.
iii) Rubber tubing and rubber stoppers
iv) Discarded cultures and contaminated media
prior to washing.
v) Laboratory aprons coats.
vi) Canned foods.
|
Only when steam sterilization is not
applicable
|
Heating
with bacteriostatic
agents, 1000C for 30 mins
|
Aqueous
preparation unsuitable for higher temperature.
i) Injection fluid
ii) Eye drops
|
i) Intrathecal, intra-atrial, peridural,
peritoneal, intra-cysternal, intracardiac, intraocular injection.
ii) Large volume parenteral
|
Tyndallization 1000C, 20 mins, 3 consecutive days,
Arnold sterilizer
|
Gelatin, milk and carbohydrate
|
|
Pasteurization
Holder method: 63 to 660C for 30 mins or
Flash method: 720C for 20 sec.
|
Beer and milk.
|
|
Dry
heat sterilization
1600C for 2 hours
|
i) Heat stable nonaqueous products and
powders.
ii) Glass wares- test tubes, pipettes,
Petri-dishes and flasks.
iii) Forceps, scalpels, scissors, throat swabs
and glass syringes.
iv) Fats, oils, greasy materials (e.g.
petroleum jelly)
|
i) Aqueous solution
ii) Surgical dressings
|
IR
radiation
1800C or 2000C
under vacuum.
|
Metal and surgical instruments Glass
syringes.
|
Same as dry heat sterilization
|
Ionizing
radiation (gamma radiation from Co-60 or Cs-137 or energized
electrons from electron accelerator.), Dose = 2.5 Mrads
|
i) Pre-packed disposable plastic syringes,
catheters, plastic tubes in saline sets etc.
ii) Packaging material made of Aluminium and
plastic strips.
|
Aqueous solution
|
UV-radiation
253.7 nm Source: Mercury vapor lamp.
|
Air sterilization, surface sterilization
|
|
Filtration
0.22
nm pore size
|
i) Serum, physiological salt solution
containing NaHCO3
ii) Enzymes, bacterial toxins
iii) Solutions of antibiotics
|
Suspensions and viscous solutions, oils
etc.
|
Gas
sterilization
|
Disposable syringes, hypodermic needles,
pre-packed materials etc
|